Display ecg in semi-auto mode, Display heart rate in semi-auto mode – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 31

Semi-Auto Mode Parameters
9650-1201-01 Rev. M
25
Display ECG in Semi-Auto Mode
This parameter allows you to set the unit to either display or not display the patient’s ECG trace
on the monitor while in Semi-Auto mode.
Values are as follows:
Display Heart Rate in Semi-Auto Mode
This parameter allows you to set the unit to display the patient heart rate on the monitor while
in Semi-Auto mode.
Values are as follows:
Auto Analyze on Power Up (Semi-Automatic Mode only)
This parameter allows you to specify whether or not the unit automatically begins an ECG
rhythm analysis when the device is turned on and pads are properly connected. When this
parameter is set to Yes, the unit automatically begins the ECG rhythm analysis. The
ANALYZING ECG message is displayed for 5 seconds and a STAND CLEAR prompt is
displayed and announced, indicating the start of the analysis. If the Auto Analyze option is set
to Yes, the analysis is the first of a stacked shock sequence.
Note:
The unit must be turned off for 10 seconds or more then turned on again for the
analysis to auto-start.
Values are as follows:
3.
Use the softkey as follows:
•
Prev Digit — moves the highlight left to the
previous digit.
•
Inc Digit or Dec Digit — increments or
decrements the value of the selected digit.
•
Next Digit — moves the highlight right to the
next digit to be entered.
•
Enter— selects the values that you entered
and returns the unit to the Semi-Auto mode
configuration menu.
Step
Action
Possible Values:
Yes, No
Default:
Yes
Possible Values:
Yes, No
Default:
Yes
Possible Values:
Yes, No
Default:
No