beautypg.com

Auto transmit after 12 lead, 12 lead fax format – ZOLL E Series Monitor Defibrillator Rev M User Manual

Page 37

background image

12-Lead Parameters

9650-1201-01 Rev. M

31

Auto Transmit After 12 Lead

When set to Yes, this parameter allows you to set the unit to automatically enter the
transmission set-up screen following completion of a 12-lead analysis.

Note:

To fax using 2x6 format, this must be set to Yes.

Values are as follows:

12 Lead Fax Format

This parameter allows you to specify the 12-lead fax format.

When set to 4x3 Standard, the 12-lead format on the fax output is identical to the format on the
strip chart with the addition of a 10-second Lead II rhythm strip. If 12 Lead Printout is set to
Cabrera, the fax is Cabrera as well.

When set to 2x6, the 12-lead format on the fax output is configured for two columns of 6 leads,
each with 5 seconds of data. The fax does not contain the 12SL interpretative statements or the
10-second Lead II rhythm strip. In addition, you must set Auto Transmit After 12 Lead to Yes.

If you select 4x3 Median Compl, the 12-lead ECG waveforms on the strip chart, summary
report, and fax will contain a single median complex beat computed by the GE Medical
Systems Information Technologies 12SL analysis program.

Values are as follows:

Possible Values:

Yes, No

Default:

No

Possible Values:

4x3 Standard, 2x6, 4x3 Median Compl

Default:

4x3 Standard