CPR Compression Metronome
This parameter allows you to select in which modes you want the CPR metronome enabled,
Semi-Auto Only (default), Yes (Manual and Semi-Auto), or No (no prompts regardless of
mode). If this parameter is set to Yes and the unit is in Manual Mode, the metronome is
activated when compressions fall to below 80 compressions per minute during CPR. If this
parameter is set to Semi-Auto Only or Yes and the unit is in Semi-Auto Mode, the metronome
is always on during CPR periods. If this parameter is set to No, the metronome is disabled.
Values are as follows:
Target CPR Depth
This parameter allows you to select the minimum compression depth required for Good
Compressions: 1.5 inches (3.8 cm) or 2.0 inches (5.0 cm). If this parameter is set to 1.5 inches,
the references markers on the CPR bar graph are displayed at 1.5 and 2.0 inches. If this
parameter is set to 2.0 inches, the reference markers on the CPR bar graph are displayed at 2.0
and 2.4 inches.
Values are as follows:
Display Fully Release Prompt
This parameter, when set to Yes, specifies that the FULLY RELEASE prompt will be displayed
every 45 seconds during CPR. If this parameter is set to No, the prompt is not displayed.
Values are as follows:
Possible Values:
Yes, Semi-Auto Only, No
Default:
Semi-Auto Only
Possible Values:
1.5 inches, 2.0 inches
Default:
1.5 inches
Possible Values:
Yes, No
Default:
No