Basic Defibrillation Parameters

9650-1201-01 Rev. M

15

Basic Defibrillation Parameters

The Basic Defibrillation parameters control functions related to performing defibrillation with
the E Series unit in manual mode and include:

Defib Default to Pads

Basic Auto Energy Escalation

Retain Sync After Defib

Defib Default to Pads

This parameter allows you to override the “Display Lead on Power-Up” setting and use Pads/
Paddles whenever the unit is switched to DEFIB mode.

Values are as follows:

Basic Auto Energy Escalation

This parameter allows you to configure the unit to automatically increment defibrillation
energy to pre-specified levels after each of the first two shocks in manual mode (see “Energy
Level: Shock 1” on page 16, “Energy Level: Shock 2” on page 16, and “Energy Level: Shock
3” on page 17). When set to Yes, the ENERGY INCREMENTED message is displayed
whenever the unit automatically increments the selected defibrillator energy after delivery of a
shock with paddles or MFE pads.

Note:

Manually changing the energy level outside the pre-programmed sequence and
delivering a shock disables the auto energy escalation function until the unit is turned
off for more than 10 seconds, then turned back on.

This option increments the energy level regardless of Shock Advisory activation or status.

Values are as follows:

Retain Sync After Defib

This parameter allows you to configure the unit to remain in SYNC mode after a synchronized
cardioversion shock. SYNC mode remains on until the Sync On/Off softkey is pressed again or
the unit is switched out of DEFIB mode.

Values are as follows:

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

No