Device Identifier
The Device Identifier feature allows you to set an 11-digit device identifier code. This code is
divided into two sub codes: the site ID (consisting of eight alphanumeric characters) and the
device ID (consisting of three alphanumeric characters). The device identifier is printed on the
12-lead strip chart, summary report, fax and Catalyst MUSE printouts.
If you are transmitting 12-lead ECG data to a Catalyst MUSE system, you should set up the
device identifier.
Values are as follows:
Procedure
Follow these steps to set up a device identifier.
4.
Use the softkeys as follows:
•
Prev Item and Next Item — scrolls the
highlight to the next or previously highlighted
code marker feature.
•
Change Value— allows you to pick a new
marker for the selected item number.
•
Delete to End — delete all configured
markers from the highlighted item number to
the end of the list. You can then enter new
clinical actions for each item.
•
Return — returns you to the previous menu.
Step
Action
Possible Values:
0 through 9, A through Z, blank and underscore inclusive for each digit.
Default:
00000000 000
Step
Action
1.
Press the Change Config softkey from the Main
System Configuration screen.
The Parameter Categories screen appears.
Select General, then press the Enter softkey.
2.
Select Device Identifier and press the Change
Value softkey.
The Device Identifier screen is displayed.
Enter
Return
Next
Item
Prev
Item
System Configuration
1 of 7
ECG
Tones
Set/Review
General
Set/Review
Set/Review
Prev
Letter
Next
Letter
Enter
Dec
Letter
Inc
Letter
System Configuration
Set Site ID
0 0 0 0 0 0 0 0
Set Device ID
0 0 0