Record spo, Record etco, Allow card erase – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 56

Chapter 2
Configuration Parameters
50
9650-1201-01 Rev. M
Record SpO
2
in Advisory Mode
This parameter allows you to set the unit to record the SpO
2
waveform continuously to the
PCMCIA card while in Manual/Advisory mode.
Values are as follows:
Record EtCO
2
in Advisory Mode
This parameter allows you to set the unit to record the EtCO
2
waveform continuously to the
PCMCIA card while in Manual/Advisory mode.
Values are as follows:
Report Memory Card Errors
This parameter allows you to disable or enable memory card error messages and voice prompts.
If this option is set to No, the unit suppresses all error messages and prompts related to the use
of the memory card (i.e., INSERT CARD, REPLACE CARD, CARD FULL, CARD LOW < n
MINUTES, CARD WRITE PROTECTED, NOT AN E SERIES DATA CARD, ERASE CARD,
and CARD NOT ERASED PROPERLY).
Values are as follows:
Allow Card Erase
This parameter, when set to Yes, allows you to erase PCMCIA data cards on the E Series unit.
Values are as follows:
Possible Values:
Yes, No
Default:
No
Possible Values:
Yes, No
Default:
No
Possible Values:
Yes, No
Default:
Yes
Possible Values:
Yes, No
Default:
Yes