Chapter 2
Configuration Parameters
42
9650-1201-01 Rev. M
Trending Parameters
Trending parameters are associated with the trend report function and how trend reports are
displayed, and include:
•
Trending Enabled
•
Trending Report Restart Delay
•
Trending Report Zoom Level
Trending Enabled
This parameter allows you to specify whether or not the E Series unit prints and displays trend
reports (as long as the appropriate hardware is installed).
Values are as follows:
Trending Report Restart Delay
This parameter allows you to review or change the length of time the device must be turned off
before it erases all data stored in trend memory.
Values are as follows:
Trending Report Zoom Level
This parameter allows you to set the power-up resolution for the displayed trend report.
Trend data values for all physiological parameters are simultaneously sampled by the E Series
and stored in memory once every 30 seconds. When displaying this data in tabular format on
the E Series screen, however, you can choose to view all or only a subset of the stored trend
data. The factory default zoom level causes all 30-second data samples to be displayed when
viewing the trend table. If lower resolution is desired, you can set this configuration option to
display only the samples taken every 5, 10, 15, 30, or 60 minutes. All data samples are printed
on the strip chart regardless of this setting.
Values are as follows:
Possible Values:
Yes, No
Default:
Yes
Possible Values:
5, 15, 30, 90 minutes, 6 hours, 12 hours, 18 hours, 1.5 days
Default:
15 minutes
Possible Values:
All (30 sec), 5 min, 10 min, 15 min, 30 min, 60 min
Default:
All