Chapter 2
Configuration Parameters
30
9650-1201-01 Rev. M
Values are as follows:
Refer to the 12- Lead ECG Monitoring Operator’s Guide insert for details on each setting.
Print 2 Copies 12 Lead
This parameter allows you to specify whether or not to print 2 copies of the 12-lead 4x3 report.
When this parameter is set to Yes, two copies of the 12-lead 4x3 report and 12SL™ analysis
print on the strip chart. This setting only affects 12-lead reports that are generated by pressing
the Acquire softkey.
Values are as follows:
Print 12 Lead Measurements
This parameter allows you to set the unit to print lead measurements. When set to Yes, the 12SL
analysis produces and prints a measurements matrix including measurements for each lead.
Global measurements are produced and printed regardless of this setting. This setting affects
the strip chart and summary report printouts as well as the fax transmission. When set to Yes,
any faxed 12-lead report contains two pages. The second page of the report contains the
computed measurements matrix. When set to No, lead measurements are not printed.
Values are as follows:
Print 12 Lead Interpretation
This parameter allows you to set the unit to print interpretive statements.
When set to Yes, the 12SL interpretive statements print following the acquisition of a
12-lead ECG. This setting affects the strip chart and summary report printouts as well as the fax
transmission.
Values are as follows:
Possible Values:
0.05-150 Hz 4x3, 0.05-150 Hz Cont, 0.05-40 Hz 4x3
Default:
0.05-40 Hz 4x3
Possible Values:
Yes, No
Default:
No
Possible Values:
Yes, No
Default:
No
Possible Values:
Yes, No
Default:
Yes