Custom 5 Wire Grouping in Use

This parameter allows you to specify whether you are using a standard (default) ECG lead
group or a custom 5 wire grouping (II, III, aVF). When this parameter value is set to No, leads
are printed as follows:

When this parameter is set to Yes, leads are printed as follows:

Values are as follows:

Selected Lead

Leads Printed (top to bottom)

I

I, II, III

II

II, I, III

III

III, I, II

aVR

aVR, aVL, aVF

aVL

aVL, aVR, aVF

aVF

aVF, aVR, aVL

V1 (5 Wire Cable)

V1, I, aVR

Selected Lead

Leads Printed (top to bottom)

II

II, III, aVF

III

III, II, aVF

aVF

aVF, II, III

Possible Values:

Yes, No

Default:

No