Restoring Default Values

9650-1201-01 Rev. M

5

Restoring Default Values

You can restore all configuration parameter values to their factory default settings. Only the
display language, if different from the default, does not change.

Procedure

Follow these steps to restore all configuration parameters to default values:

If you want to review configuration settings, press the Change Config softkey. See “Accessing
and Changing Configuration Parameters” on page 6 for more information. If you want to leave
System Configuration mode, press the Exit Config softkey. See “Leaving Configuration
Mode” on page 6 for more information.

Step

Action

1.

Power on the unit in System Configuration mode
as described in “Entering System Configuration
Mode” on page 2.

2.

Press the Restore Defaults softkey to
reconfigure parameter values to their default
settings.

Once default settings are restored, the unit
displays the DEFAULTS RESTORED and
PRESS CHANGE CONFIG TO VIEW messages.

0

3


Delete

2

1

System Configuration

Enter Configuration Passcode

or Power Off to Exi

t

? ? ? ? ? ? ? ?

Print

Config

Change

Config

Exit

Config

Read
From

Card

Restore

Defaults

System Configuration

Defaults Restored

Press Change Config to View