Chapter 2

Configuration Parameters

52

9650-1201-01 Rev. M

The unit has delivered a pacing current greater than 25 mA since power-up.

When this option is enabled, the Remove Card prompt is displayed for 10 seconds, then the
E Series unit shuts off.

Values are as follows:

Sync Defib Excluded

This parameter allows the exclusion of synchronized cardioversion from the criteria for
determining whether the Erase Card or Remove Card messages are issued following power off.

When set to Yes, if the E Series unit has only discharged self-test or synchronized cardioversion
defibrillations, and has not delivered pacing current greater than 25 mA, the unit displays the
Erase Card -- Yes/No? prompt at power off if so configured. Otherwise, the unit displays the
Remove Card prompt at power off if so configured.

Values are as follows:

Possible Values:

Yes, No

Default:

No

Possible Values:

Yes, No

Default:

No