12-Lead Parameters

9650-1201-01 Rev. M

29

12-Lead Parameters

12-Lead configuration parameters are only available if the 12-Lead ECG monitoring option is
installed in the unit, and include:

Print 12 Lead 4 X 3

12 Lead Printout

12 Lead Freq Response

Print 2 Copies 12 Lead

Print 12 Lead Measurements

Print 12 Lead Interpretation

Auto Transmit After 12 Lead

12 Lead Fax Format

Set Lead Groups

Lead Group Default

12 Lead Secs Per Lead

12 Lead Print Speed

PTCA Message Enabled

PTCA Threshold

Print 12 Lead 4 X 3

This parameter value indicates the number of copies of a 12-lead 4x3 report (or limited 2 x 3
report) printed after pressing and holding the RECORDER button for three seconds. When set
to None, pressing and holding the RECORDER button causes a one- or three-lead DIAG
BANDWIDTH ECG strip to print.

Values are as follows:

12 Lead Printout

This parameter allows you to specify the format for 12-lead 4x3 printouts (or limited 2 x 3
reports). When set to Standard, the 12-lead 4x3 printout uses the standard lead format. When
set to Cabrera, the 12-lead 4x3 printout uses the Cabrera-style lead format.

Values are as follows:

12 Lead Freq Response

This parameter allows you to set the 12-lead filter settings and printing format.

Possible Values:

1 Copy, 2 Copies, None

Default:

None

Possible Values:

Standard, Cabrera

Default:

Standard