Energy level: shock 3, Auto charge in advisory mode, Auto analyze – ZOLL E Series Monitor Defibrillator Rev M User Manual

Page 23

background image

Advisory Defibrillation Parameters

9650-1201-01 Rev. M

17

Values are as follows:

Energy Level: Shock 3

This parameter allows you to select the energy level used for the third and all subsequent
shocks delivered by the defibrillator when the unit is:

in semi-automatic or advisory mode and using pads

in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes

Values are as follows:

Auto Charge in Advisory Mode

This parameter allows you to specify whether the unit automatically charges the defibrillator to
the configured or user-selected energy setting after detection of a shockable rhythm by ECG
analysis. When set to Yes, the unit automatically charges the defibrillator to the configured or
user-selected energy setting. When set to No, it does not.

Values are as follows:

Auto Analyze

This parameter allows you to specify whether the unit automatically repeats ECG rhythm
analyses after the delivery of shocks in the AED and Advisory modes and works in conjunction
with the Stacked Shocks parameter described next. Enabling this parameters also enables the
occurrence of post-shock CPR intervals. When set to No, which is the Manual mode default,
the unit does not perform foreground analysis automatically (with the exception of the first
analysis on power-up in Auto-Defib mode, if the unit is configured to analyze on power-up).

When set to Yes, and the Stacked Shocks options is set to 3, the unit automatically analyzes the
patient’s ECG rhythm after the first and second shocks in each three-shock sequence.
Following delivery of the final shock, the unit stops automatic re-analysis and either displays a
CHECK PULSE prompt, or immediately begins a CPR interval, depending on additional
parameter settings. This function operates only when the Auto Charge in Advisory Mode is set
to Yes.

When set to Semi-Auto Only, the unit functions as described above only when it is in AED
Semi-Auto mode. If the unit is changed to Manual mode from AED Semi-Auto mode, the unit

Possible Values:

1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J

Default:

150 J

Possible Values:

1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J

Default:

200 J

Possible Values:

Yes, No

Default:

Yes