CPR Parameters

The CPR parameters are attributes that relate to the E Series Real CPR Help

®

functionality and

include:

Display CPR Compression Bar Graph

CPR Record Data to Card

CPR Guidance Audio Prompts

CPR Compression Metronome

Target CPR Depth

Display Fully Release Prompt

Display CPR Compression Bar Graph

When set to Yes, this parameter specifies that the CPR compression bar graph will
automatically replace the second waveform on the E Series display whenever chest
compressions are detected.

Values are as follows:

CPR Record Data to Card

This parameter, when set to Yes, causes CPR data to be recorded to the PCMCIA card

whenever CPR-Dpadz

®

are in use. This parameter can be set to Yes only if Record ECG Data

in Advisory Mode is set to Yes.

Values are as follows:

CPR Guidance Audio Prompts

This parameter allows you to select the modes in which CPR-related audio prompts will be
issued, Semi-Auto Only (default), Yes (Manual and Semi-Auto), or No (no prompts regardless
of mode). In Semi-Auto Only, or Yes, the audio prompts are:

PUSH HARDER when the detected compression depth is consistently less than the target
CPR depth.

GOOD COMPRESSIONS when

the rescuer respon

ds by increasing compression depth to

the target CPR depth or more.

Values are as follows:

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

No

Possible Values:

Yes, Semi-Auto Only, No

Default:

Semi-Auto Only