General parameters, Procedure, Code markers – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 14: Device identifier, Configuration mode passcode, System configuration

Chapter 2
Configuration Parameters
8
9650-1201-01 Rev. M
General Parameters
General parameters are attributes that affect the unit’s basic operation and accessibility and
include:
•
Code Markers
•
Device Identifier
•
Configuration Mode Passcode
Code Markers
This parameter allows you to review and change the preconfigured list of Code Markers for use
in DEFIB, MONITOR, and PACE modes. Code Marker settings are language dependent. You
can specify up to three groups (MONITOR, DEFIB, and PACE) of 20 entries each.
Note:
MONITOR code markers are not used in AED units.
Procedure
Follow these steps to review, change, or reset code marker settings:
Step
Action
1.
Press the Change Config softkey from the Main
System Configuration screen.
The Parameter Categories screen appears.
2.
Select General and press the Enter softkey, then
press the Change Value softkey.
3.
Select either DEFIB CODE MARKERS,
MONITOR CODE MARKERS, or PACE CODE
MARKERS, then press the Change Value
softkey to toggle between Change/Review or Set
to Defaults.
•
If you select Set to Defaults, press the Enter
softkey to set the chosen code markers to
their default values.
•
If you select Change/Review, press the Enter
softkey to access the screen that allows you
to change the marker, as shown.
This example shows the DEFIB CODE
MARKERS screen.
Enter
Return
Next
Item
Prev
Item
System Configuration
1 of 7
ECG
Tones
Set/Review
General
Set/Review
Set/Review
Delete
to End
Return
Next
Item
Prev
Item
System Configuration
Item 2: Intubate
Item 3: EPI
CPR
Item 1: CPR
Defib Markers
Markers List
Item 4: LIDO
Item 5: Narcan
Item 6: B-block
Item 7: Atrovent
Item 8: Mag Sulf
EPI
Atrop
DOPA
PHEN
Bicarb
Aspirin
Oxygen
Change
Value