beautypg.com

General parameters, Procedure, Code markers – ZOLL E Series Monitor Defibrillator Rev M User Manual

Page 14: Device identifier, Configuration mode passcode, System configuration

background image

Chapter 2

Configuration Parameters

8

9650-1201-01 Rev. M

General Parameters

General parameters are attributes that affect the unit’s basic operation and accessibility and
include:

Code Markers

Device Identifier

Configuration Mode Passcode

Code Markers

This parameter allows you to review and change the preconfigured list of Code Markers for use
in DEFIB, MONITOR, and PACE modes. Code Marker settings are language dependent. You
can specify up to three groups (MONITOR, DEFIB, and PACE) of 20 entries each.

Note:

MONITOR code markers are not used in AED units.

Procedure

Follow these steps to review, change, or reset code marker settings:

Step

Action

1.

Press the Change Config softkey from the Main
System Configuration screen.

The Parameter Categories screen appears.

2.

Select General and press the Enter softkey, then
press the Change Value softkey.

3.

Select either DEFIB CODE MARKERS,
MONITOR CODE MARKERS, or PACE CODE
MARKERS, then press the Change Value
softkey to toggle between Change/Review or Set
to Defaults.

If you select Set to Defaults, press the Enter
softkey to set the chosen code markers to
their default values.

If you select Change/Review, press the Enter
softkey to access the screen that allows you
to change the marker, as shown.

This example shows the DEFIB CODE
MARKERS screen.

Enter

Return

Next
Item

Prev

Item

System Configuration

1 of 7

ECG

Tones

Set/Review

General

Set/Review

Set/Review

Delete

to End

Return

Next

Item

Prev

Item

System Configuration

Item 2: Intubate

Item 3: EPI

CPR

Item 1: CPR

Defib Markers

Markers List

Item 4: LIDO

Item 5: Narcan

Item 6: B-block

Item 7: Atrovent

Item 8: Mag Sulf

EPI

Atrop

DOPA

PHEN

Bicarb

Aspirin

Oxygen

Change

Value