Chapter 2

Configuration Parameters

40

9650-1201-01 Rev. M

Procedure

Follow these steps to change the state or value of any of the alarms:

Systolic Blood
Pressure

Default: 160 mmHg or 21.3 kPa
Possible Values: In mmHg: 145 -
260 mmHg inclusive in increments
of 5

In kPa: 12.7, 13.3, 14.0, 14.7, 15.3,
16.0, 16.7, 17.3, 18.0, 18.7, 19.3,
20.0, 20.7, 21.3, 22.0, 22.7, 23.3,
24.0, 24.7, 25.3, 26.0, 26.7, 27.3,
28.0, 28.7, 29.3, 30.0, 30.7, 31.3,
32.0, 32.7, 33.3, 34.0, 34.7

Default: 90 mmHg or 12.0 kPa
Possible Values: In mmHg: 40 - 140
mmHg inclusive in increments
of 5

In kPa: 5.3, 6.0, 6.7, 7.3, 8.0, 8.7,
9.3, 10.0, 10.7, 11.3, 12.0, 12.7, 13.3,
14.0, 14.7, 15.3, 16.0, 16.7, 17.3,
18.0, 18.7

Diastolic Blood
Pressure

Default: 110 mmHg or 14.7 kPa
Values: In mmHg: 105 - 200
inclusive in increments of 5

In kPa: 14.0, 14.7, 15.3, 16.0, 16.7,
17.3, 18.0, 18.7, 19.3, 20.0, 20.7,
21.3, 22.0, 22.7, 23.3, 24.0, 24.7,
25.3, 26.0, 26.7

Default: 50 mmHg or 6.7 kPa
Values: In mmHg: 25 - 100 inclusive
in increments of 5

In kPa: 3.3, 4.0, 4.7, 5.3, 6.0, 6.7,
7.3, 8.0, 8.7, 9.3, 10.0, 10.7, 11.3,
12.0, 12.7, 13.3

Mean Blood
Pressure

Default: 130 mmHg or 17.3 kPa
Values: In mmHg: 125 - 220
inclusive in increments of 5

In kPa: 16.7, 17.3, 18.0, 18.7, 19.3,
20.0, 20.7, 21.3, 22.0, 22.7, 23.3,
24.0, 24.7, 25.3, 26.0, 26.7, 27.3,
28.0, 28.7, 29.3

Default: 60 mmHg or 8.0 kPa
Values: In mmHg: 30 -120 inclusive
in increments of 5

In kPa: 4.0, 4.7, 5.3, 6.0, 6.7, 7.3,
8.0, 8.7, 9.3, 10.0, 10.7, 11.3, 12.0,
12.7, 13.3, 14.0, 14.7, 15.3, 16.0

Step

Action

1.

Press the Change Config softkey from the Main
System Configuration screen.

The Parameter Categories screen appears.

Select Alarms, then press the Enter softkey.

2.

Select Alarm Limits at Power Up, and press the
Change Value softkey.

The Set Alarm Defaults screen is displayed.

Alarm

Maximum Default/Values

Minimum Default/Values

Enter

Return

Next

Item

Prev

Item

System Configuration

4 of 7

NIBP

Trending

Set/Review

Alarms

Set/Review

Set/Review

Select
Param

Next

Field


Return

Dec

<

Inc

>

System Configuration

SET ALARM DEFAULTS

Parameter

State

Low

High

SYSTOLIC
DIASTOLIC
MEAN
MORE

ENABLE
ENABLE
ENABLE

30
90
50
60

150
160
110
130

ECG HR

ENABLE