Procedure – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 46

Chapter 2
Configuration Parameters
40
9650-1201-01 Rev. M
Procedure
Follow these steps to change the state or value of any of the alarms:
Systolic Blood
Pressure
Default: 160 mmHg or 21.3 kPa
Possible Values: In mmHg: 145 -
260 mmHg inclusive in increments
of 5
In kPa: 12.7, 13.3, 14.0, 14.7, 15.3,
16.0, 16.7, 17.3, 18.0, 18.7, 19.3,
20.0, 20.7, 21.3, 22.0, 22.7, 23.3,
24.0, 24.7, 25.3, 26.0, 26.7, 27.3,
28.0, 28.7, 29.3, 30.0, 30.7, 31.3,
32.0, 32.7, 33.3, 34.0, 34.7
Default: 90 mmHg or 12.0 kPa
Possible Values: In mmHg: 40 - 140
mmHg inclusive in increments
of 5
In kPa: 5.3, 6.0, 6.7, 7.3, 8.0, 8.7,
9.3, 10.0, 10.7, 11.3, 12.0, 12.7, 13.3,
14.0, 14.7, 15.3, 16.0, 16.7, 17.3,
18.0, 18.7
Diastolic Blood
Pressure
Default: 110 mmHg or 14.7 kPa
Values: In mmHg: 105 - 200
inclusive in increments of 5
In kPa: 14.0, 14.7, 15.3, 16.0, 16.7,
17.3, 18.0, 18.7, 19.3, 20.0, 20.7,
21.3, 22.0, 22.7, 23.3, 24.0, 24.7,
25.3, 26.0, 26.7
Default: 50 mmHg or 6.7 kPa
Values: In mmHg: 25 - 100 inclusive
in increments of 5
In kPa: 3.3, 4.0, 4.7, 5.3, 6.0, 6.7,
7.3, 8.0, 8.7, 9.3, 10.0, 10.7, 11.3,
12.0, 12.7, 13.3
Mean Blood
Pressure
Default: 130 mmHg or 17.3 kPa
Values: In mmHg: 125 - 220
inclusive in increments of 5
In kPa: 16.7, 17.3, 18.0, 18.7, 19.3,
20.0, 20.7, 21.3, 22.0, 22.7, 23.3,
24.0, 24.7, 25.3, 26.0, 26.7, 27.3,
28.0, 28.7, 29.3
Default: 60 mmHg or 8.0 kPa
Values: In mmHg: 30 -120 inclusive
in increments of 5
In kPa: 4.0, 4.7, 5.3, 6.0, 6.7, 7.3,
8.0, 8.7, 9.3, 10.0, 10.7, 11.3, 12.0,
12.7, 13.3, 14.0, 14.7, 15.3, 16.0
Step
Action
1.
Press the Change Config softkey from the Main
System Configuration screen.
The Parameter Categories screen appears.
Select Alarms, then press the Enter softkey.
2.
Select Alarm Limits at Power Up, and press the
Change Value softkey.
The Set Alarm Defaults screen is displayed.
Alarm
Maximum Default/Values
Minimum Default/Values
Enter
Return
Next
Item
Prev
Item
System Configuration
4 of 7
NIBP
Trending
Set/Review
Alarms
Set/Review
Set/Review
Select
Param
Next
Field
Return
Dec
<
Inc
>
System Configuration
SET ALARM DEFAULTS
Parameter
State
Low
High
SYSTOLIC
DIASTOLIC
MEAN
MORE
ENABLE
ENABLE
ENABLE
30
90
50
60
150
160
110
130
ECG HR
ENABLE