Advisory Defibrillation Parameters

9650-1201-01 Rev. M

19

At the end of a stacked shock sequence (assuming the final shock is delivered) and when
the optional 10 second CHECK PULSE message disappears, the CPR interval begins.
The configured CPR prompt is displayed for the duration of the CPR interval.

If the defibrillator disarms itself after it was charged (because the Defib Ready time
elapsed before a shock was delivered), the unit displays the PRESS ANALYZE prompt
until the ANALYZE button is pressed.

The Display DO CPR Message feature is operational only when the Auto Charge in Advisory
Mode and Auto Analyze options are both set to Yes.

Note:

When set to Only in Semi-Auto, the unit functions as described above only when it is
in AED semi-auto mode.

Values are as follows:

CPR Message After Shock

This parameter determines which CPR message/voice prompt is issued during the CPR interval
after the delivery of the final shock in a stacked shock sequence. The selected message is
displayed throughout the entire CPR interval.

Values are as follows:

CPR Message After No Shock Advised

This parameter determines which CPR message/voice prompt is issued during the CPR interval
after a No Shock Advised analysis result. The selected message is displayed throughout the
entire CPR interval.

Values are as follows:

Duration of CPR Interval

This parameter allows you to specify the duration of the CPR interval as described in the
Display DO CPR Message parameter. Values are as follows:

Possible Values:

Yes, No, Semi-Auto Only

Default:

Manual/Advisory units: No, AED units: Semi-Auto Only

Possible Values:

IF NO PULSE, PERFORM CPR; PERFORM CPR

Default:

PERFORM CPR

Possible Values:

IF NO PULSE, PERFORM CPR; PERFORM CPR

Default:

IF NO PULSE, PERFORM CPR

Possible Values:

1, 1.5, 2, 2.5, 3, 3.5 or 4 minutes

Default:

2 minutes