Valid values are as follows:

Enable RescueNet Link

This options allows the E Series unit to automatically communicate with RescueNet Link host
software via a Bluetooth connection. If there is no valid host, the on-screen realtime mode
indicator is displayed as green, and the unit will keep track of 12-lead patient records or
trending data until it re-enters the range of the receiving RescueNet Link host. The data is then
automatically transmitted. Following a defibrillation discharge, automatic data transmission
will be suspended momentarily and will resume automatically if it was enabled prior to the
discharge.

With the RescueNet Link option enabled, the automatic data transmission port cannot be
changed from ZDR Mobile. Patient records cannot be transmitted to configured ZDR Server
contacts using Dial-Up Networking.

Automatic data transmission can be deactivated and re-activated (see Appendix B of the
E Series Operator’s Guide for more information about automatic data transmission).

Values are as follows:

Recipient Contact List

This parameter allows you to manage the list of email or FAX recipients of 12-lead reports.

Four different types of recipients can be added to the list. A FAX phone number, e-mail
address, or distribution list all go to their intended recipients via the ZOLL Data Relay server. It
is also possible to enter a FAX phone number for sending in Direct FAX mode (via modem).
Up to 36 contacts can be added to the Recipient Contact List.

For instructions on how to set up a distribution list, see the ZOLL Data Relay System
Configuration Guide.

Procedure

To add recipients to the Recipient Contact List:

Possible Values:

Yes

,

No

Default:

Yes

Possible Values:

Yes, No

Default:

No