Ecg monitoring parameters, Notch filter, Frequency response – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 18: Display lead on power-up

Chapter 2
Configuration Parameters
12
9650-1201-01 Rev. M
ECG Monitoring Parameters
ECG monitoring parameters are attributes associated with using the E Series unit for ECG
monitoring, and include:
•
Notch Filter
•
Frequency Response
•
Display Lead on Power-Up
•
Gain on Power-Up
•
Enable Leads Off
•
Enable Pacer Detection
Notch Filter
This parameter allows you to set the ECG Notch Filter frequency for proper ac mains
interference rejection.
Values are as follows:
Frequency Response
This parameter allows you to select the ECG filter bandwidth to be used during ECG
monitoring. You can set the unit’s frequency to either help reduce noise, help with the
diagnostic response, or set it as a default response.
Values are as follows:
Display Lead on Power-Up
This parameter allows you to select the lead to be displayed on power-up.
For AED units, when the unit powers on in semi-automatic mode, this option is ignored. The
unit powers on in Lead II, and switches leads automatically between leads II and MFE for Auto
Monitor and Auto Defib modes, respectively. When the unit enters Manual Mode (Manual
Mode override sequence is completed), the lead switches to the setting specified in the “Display
Lead on Power-Up” option.
Values are as follows:
Possible Values:
50 Hz, 60 Hz
Default:
60 Hz
Possible Values:
0.67 - 21 Hz, 0.67 - 27 Hz, 1 - 21 Hz
Default:
0.67 - 21 Hz
Possible Values:
Paddles/Pads, Lead I, Lead II, Lead III
Default:
Paddles/Pads