Pace Parameters

The Pace parameters are attributes that control functions related to performing pacing with an
E Series unit and include:

Pace Rate Setting on Power Up

Async Softkey in Pace Mode

Pace Rate Setting on Power Up

This parameter allows you to set the default pace rate, pulses per minute (ppm), of the E Series
unit at power-on.

Values are as follows:

Async Softkey in Pace Mode

This parameter allows you to set the unit to display the Async Pacing On/Off softkey when the
unit is in pace mode. For information about Asynchronous Pacing, refer to the E Series
Operator’s Guide.

Values are as follows:

Possible Values:

30 ppm to 102 ppm (inclusive in increments of 4 ppm)

Default:

70 ppm

Possible Values:

Yes, No

Default:

Yes