Nibp auto interval default, Nibp first inflate pressure default, Nibp button press and hold – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 44

Chapter 2
Configuration Parameters
38
9650-1201-01 Rev. M
NIBP Auto Interval Default
This parameter allows you to set the default for the amount of time between automatic
measurements.
Values are as follows:
NIBP First Inflate Pressure Default
This parameter allows you to set the default NIBP cuff inflation pressure (the pressure to which
the NIBP cuff inflates at the beginning of its first measurement after power up).
Values are as follows:
NIBP Button Press and Hold
This parameter allows you to define which NIBP function is activated when you press and hold
the front panel NIBP button. You can configure the unit to either:
•
Begin a sequence of STAT NIBP measurements, or
•
Begin taking NIBP measurements at repeating, selected intervals.
Values are as follows:
Possible Values:
2.5, 5, 10, 15, 20, 30, 45, 60, 90, 120 minutes
Default:
30 minutes
Possible Values:
mmHg: 120, 140, 160, 180, 200, 220, 240, 260
kPa: 16.0, 18.7, 21.3, 24.0, 26.7, 29.3, 32.0, 34.7
Default:
180 mmHg
Possible Values:
NIBP Auto, NIBP Stat
Default:
NIBP Auto