9.7.2 Rotate to zero
If the
Video Watermark
was rotated, the
Rotate to zero
menu will be enabled.
Choosing this menu will rotate the
Video Watermark
to zero degree.
9.8 Gray Calibration•••
This function can make specified area image brightness a desired value among
various scenarios, achieving the continuity requirement of the observation. The
Gray
Calibration
steps are summarized as follows:
1. Click
Exposure & Gain
on the
Camera
sidebar to extend the
Exposure & Gain
group,
uncheck
Auto Exposure
box.
2. Choose
Edit>Image Select
command or click
on the toolbar to select a
reference region, and choose “
Setup>Gray Calibration
” command. A dialog called
Gray Calibration
will be brought up to display the current ROI
Average Gray
. Now the
brightness of the microscope light source can be adjusted until the
Average Gray
closing to the desired value. Click “
OK
” to finish the calibration and return to the
application area. The desired value here is around is 200.
- T510 NG Digital Camera T130 NG Digital Camera C1400 NG Digital Camera, USB 2.0 C510 NG Digital Camera, USB 2.0 T310 NG Digital Camera C310 NG Digital Camera, USB 2.0 C130 NG Digital Camera, USB 2.0 C35 NG 350K px Digital Camera, USB 2.0 D50L NG Digital Microscope D870T Digital Trinocular Microscope D670T Digital Trinocular Microscope D320L Digital Biological Microscope D2L NG Digital Microscope