3 calibrate, Calibrate – Levenhuk C800 NG Digital Camera, USB 2.0 User Manual
Page 135

Help Manual
127
should look like
1. Select the
Magnification
item, click
Delete•••
to delete the selected item;
2. Select the
Magnification
item, click the
Up
or
Down
button to modify the
Magnification
list order;
3. Click
Clear All•••
to delete all of the
Magnification
items; If one wishes to reuse the
Magnification
, a backup is needed first;
4. Click
Export•••
to back up the
Magnification
in a safe media. The file extension is
“
*.magn
”;
5. After the new installation is finished, the
Magnification
can be import by choosing
Options>Magnification•••
and click
Import•••
to load the previously saved “
*.magn
”
file;
6. If everything is ok, press
OK
to end the
Magnification
management dialog.
15.3 Calibrate•••
The detailed calibration steps are as follows:
1. Run ToupView;
2. Connect the camera to the computer and microscope;
3. Start the camera (Here, it is
UCMOS00350KPA
);
4. Switch the microscope object to 10X and put the TS-M1 stage micrometer in the
- T510 NG Digital Camera T130 NG Digital Camera C1400 NG Digital Camera, USB 2.0 C510 NG Digital Camera, USB 2.0 T310 NG Digital Camera C310 NG Digital Camera, USB 2.0 C130 NG Digital Camera, USB 2.0 C35 NG 350K px Digital Camera, USB 2.0 D50L NG Digital Microscope D870T Digital Trinocular Microscope D670T Digital Trinocular Microscope D320L Digital Biological Microscope D2L NG Digital Microscope