B&K Precision 2516 - Manual User Manual
Page 71

71
Table 20 - Video Trigger Menu 2/2
Menu
Option
Setting
Description
Type
Video
When you select type video and select
AC coupling, you could trigger to a
NTSC, PAL and SECAM video signal.
Standard
NTSC,
PAL/SECAM
Select the video standard for sync and line
number count.
Mode
Auto
Use this mode to let the acquisition free-
run in the absence of a valid trigger. This
mode allows an un-triggered, scanning
waveform at 100 ms/div or slower time
base settings.
Normal
Use this mode when you want to see only
triggered waveforms; the oscilloscope does
not display a waveform until after the first
trigger.
Single
Use this mode to run a single capture.
Setup
Enter the Video Trigger Setup Menu. See
Table 16 for details.
Next Page
Page 2/2
Return the first page of Video Trigger
Menu.
Table 21 - Video Trigger Setup Menu
Menu
Option
Setting
Description
Hold off
Adjusts the hold off time.
Hold off
Reset
Reset hold off time to 100 ns.