2.4 Perform Self-Calibration
We recommend you to perform self-calibration if any of the following
applies:
-
Operating temperature changes by more than 5 °C after
30 minutes of use.
-
Abnormal behaviors and operations are present.
-
Measurements are out of accuracy.
CAUTION:
Before performing self-calibration, disconnect all
inputs of the instrument. Otherwise, the instrument
may not calibrate correctly or may cause errors.