beautypg.com

B&K Precision 2706B - Datasheet User Manual

Manual ranging tool kit, Dmm with temperature, Model 2706b data sheet

background image

w w w.bk prec ision.c om

Tel.: 714.921.9095

DC Current

Ranges

200 uA, 20 mA, 200 mA

Resolution

0.1 uA, 10 uA, 100 uA

Accuracy

±(1.0% rdg + 1 dgts)

Input Protection

0.25 A/500 V fast blow ceramic fuse

AC Cu rrent (50 Hz -500 Hz )

Ranges

200 uA, 20 mA, 200 mA

Resolution

0.1 uA, 10 uA, 100 uA

Accuracy

±(1.5% rdg + 5 dgts)

Input Protection

0.25 A/500 V fast blow ceramic fuse

Resis tan ce

Ranges

200 Ω, 2 kΩ, 200 kΩ, 20 MΩ

Resolution

0.1 Ω, 1.0 Ω, 100 Ω, 10 kΩ,

Accuracy

200 Ω to 200 kΩ ranges: ±(1.0 % rdg + 4 dgts)
20MΩ range: ±(2.0 % rdg + 4 dgts)

Open Circuit Voltage (typical) 0.3 VDC (3.0 VDC on 200 Ω)
Overload Protection

500 VDC/ACrms

Capacitance

Ranges

200 uF, 2 mF, 20 mF

Resolution

0.1 uF, 1.0 uF, 10 uF

Accuracy

±(4 % rdg + 10 dgts)

Overload Protection

0.25 A/500 V fast blow ceramic fuse

Fre qu ency

Ranges

10 Hz to 40 kHz

Resolution

1 Hz

Accuracy

±(0.1 % rdg + 3 dgts)

Sensitivity

3.5 Vrms minimum

Overload Protection

500 VDC/AC rms

Diode Te st

Test Current

1.0 mA (typical)

Resolution

10 mV

Accuracy

±(1.5 % rdg + 3 dgts)

Open Circuit Voltage

3.0 VDC (typical)

Overload Protection

500 VDC/ACrms

Con tinuity

Audible Indication

Less than 100 Ω

Response Time

100 ms

Overload Protection

500 VDC/ACrms

Temperature

Ranges

-35ºC to 750ºC, -30ºF to 1400ºF

Resolution

0.1ºC, 0.1ºF

Accuracy

0ºC to 400ºC ±(1.0 % rdg + 1ºC)
-35ºC to 0ºC, 400ºC to 750ºC ±(3.0 % rdg + 3ºC)
-4ºF to 750ºF ±(1.0 % rdg + 2ºF)
-30ºF to -4ºF, 750ºF to 1400ºF ±(3.0 % rdg + 6ºF)

Sensor Type

K-type thermocouple

Overload Protection

60 VDC or 30 VACrms

General

Display

3 1/2 digit, 2000 count LCD

Polarity

Automatic, positive implied, negative polarity indication

Overrange

OL or -OL is displayed

Low Battery Indication

Battery symbol will display when battery voltage drops below
operating levels

Measurement Rate

2.5 times per second

Auto Power Off

Approximately 25 minutes

Operating Environment

0ºC to 50ºC at <70 % relative humidity

Storage Environment

-20ºC to 60ºC at 0 to 80 % relative humidity

Accuracies

Stated accuracy at 23ºC +5ºC at <75 % relative humidity

Temperature Coefficient

0.1 x (specified accuracy) per ºC. (ºC to 18ºC, 28ºC to 50ºC)

Power

Single standard 9 volt battery, NEDA 1604, JIS 006P, IEC 6F22

Battery Life

150 hours typical

Dimensions

165 mm (H) x 78 mm (W) x 42.5 mm (D)

Weight

Approximately 10 oz. (285 g) with battery and hanger

Manual Ranging Tool Kit

®

DMM with Temperature

Model 2706B

Data Sheet

Technical data subject to change
© B&K Precision Corp. 2013
Page 1 of 2

Ultra bright blue LCD back
light for low ambient light
measurements

LC D bac k light

The 2706B is a versatile manual-ranging DMM with temperature measurement.
The 2000 count LCD has a bright blue back light that allows the meter to be
used in dimly lit environments. With the auto-power-off feature you won’t have to
worry about accidental battery consumption. Value-packed features make this
meter a must in every "Tool Kit

®

".

Features

Auto power off

AC voltage to 750 V

DC voltage to 1000 V

Capacitance to 20 mF

Resistance to 20 MΩ

Continuity test

One-handed operation

Diode test

Magnetic hanger

Bright blue LCD backlight

D C Volts

Ranges

200 mV, 2000 mV, 20 V, 200 V 1000 V

Resolution

0.1 mV, 1 mV, 10 mV, 100 mV, 1 V

Accuracy

±(0.5 % rdg + 1 dgts)

Input Impedance

10 MΩ

Overload Protection

1000 VDC or 750 VACrms*

A C Volts (50 Hz -500 Hz )

Ranges

200 mV, 2000 mV, 20 V, 200 V 750 V

Resolution

0.1 mV, 1 mV, 10 mV, 100 mV, 1 V

Accuracy

200 mV & 20 V ranges: ±(1.2 % rdg + 5 dgts)
200 V & 750 V ranges: ±(2.0 % rdg + 5 dgts)

Input Impedance

10 MΩ

Overload Protection

1000 VDC or 750 VACrms*

*= 600 VDC/Acrms for 15 seconds on 200 mV range

Specifications

2706B

model

Supplied Accessories: User manual, one set of test leads, one spare fuse, 9V battery (installed),

and one K-type thermocouple

Three-Year Warranty