Save recorded data to external memory – B&K Precision 2516 - Manual User Manual
Page 129

129
Record Data by Manual Trigger
By default, data is recorded automatically and continuously. To record data
by a manual trigger, access the second page of the Scope Trend Plot menu
and press
to On or Off the Manual option.
When set to On, the plot will stop recording. Instead, the
button
will function as a manual trigger button. Each subsequent press will trigger
one measurement recording to the plot (both Param A and Param B).
NOTE:
In this mode, the
button will be used as a
trigger instead of a means to enter the Recorder
menu. To exit out of Scope Trend Plot, press
to select Return. All previously recorded
data will be lost if you exit. A confirmation
message will ask if you want to exit.
Save Recorded Data to External Memory
The recorded data on the scope trend plot is stored temporarily. All
recorded data will be lost if the recording gets reset, exit out of the scope
trend plot mode, or power OFF the instrument.
However, the data can be saved onto an external USB flash drive.
To save:
1.
Go to the second page of the Scope Trend Plot menu and select
CSV by pressing
.
2.
The Directory menu will be displayed.