B&K Precision 391AKIT - Datasheet User Manual
Handheld digital multimeters, Test bench, Series data sheet
High performance and value priced, the Test Bench
®
Series offers more features for the dollar than other multimeters. In addition to measuring voltage and current,
the meters also measure capacitance, frequency and temperature, and include component test, diode test and logic indicator capabilities. See the chart below for
the meter that best fits your needs. A double injection molded case provides improved grip and protection for the meters.
Features and Benefits
■
Resistance measurement
■
Frequency measurement
■
Diode check
■
Audible continuity
■
All current ranges fused
■
Auto power off
■
Ruggedized case
■
CATIII 1000 V, (all models)
■
CATIV 600 V, (392 & 393)
Handheld Digital Multimeters
Test Bench
®
Series
Data Sheet
w w w.bk prec ision.c om
Technical data subject to change
© B&K Precision Corp. 2014
High Quality, Ruggedized, Multifunctional
Features
392 & 393
391A
390A
389A
388B
True RMS
√
√
-
-
-
Ranging
Auto/Manual
Manual
Auto/Manual
Auto/Manual
Manual
DCV Accuracy
0.08%
0.05%
0.1%
0.25%
0.5%
AC/DC Voltage and Current
√
√
√
√
√
Display Digits, Count
60000, 4 5/6
20000, 4 1/2
4000, 3 3/4
4000, 3 3/4
4000, 3 3/4
Bar Graph
√
-
√
√
-
Capacitance Measurement
√
-
√
√
√
Transistor Test
-
-
-
-
√
Temperature Probe
√
-
√
-
-
Logic Probe
√
√
-
-
√
Relative Mode
√
-
√
√
-
Min/Max Hold
√
-
√
√
-
Peak Hold
√
-
√
√
-
Data Hold
√
√
√
√
-
USB
393 only
-
√
-
-
Specifications
392 & 393
391A
390A
389A
388B
Volts
True RMS reading
True RMS reading
-
-
-
DC Ranges
600 mV, 6 V, 60V, 600 V, 1000 V
200 mV, 2 V, 20 V, 200 V, 1,000 V
400 mv, 4 V, 40 V, 400 V, 1000 V
AC Ranges
600 mV, 6 V, 60V, 600 V, 750 V
200 mV, 2 V, 20 V, 200 V, 750 V
400 mv, 4 V, 40 V, 400 V, 750 V
Resolution
0.01 mV, 0.1 mV, 1 mV, 10 mV, 100 mV
10 µV, 100 mV, 1 mV 10 mV, 100 mV
100 µV, 1 mV, 10 mV, 100 mV, 1 V
Basic DC Accuracy
±(0.08% rdg + 5 dgt)
±(0.05% + 3 dgt)
±(0.1% rdg + 2 dgt)
±(0.25% rdg + 2 dgt)
±(0.5% rdg + 1 dgt)
Basic AC Accuracy
±(1.5% rdg + 20 dgt) 500 Hz -1 kHz
±(2% + 10 dgt) 500 Hz - 2 kHz
±(1.5% rdg +5 dgt) 500 Hz - 1 kHz
±(1.5% rdg +3 dgt) 500 Hz - 1 kHz
Overload Protection
1000 VDC or 750 VAC rms
1200 VDC or AC rms
500 VDC/AC rms 15 sec
on 200 mV range
1100 VDC or AC rms
1000 VDC or peak AC
1200 VDC or AC rms
500 VDC/AC rms 15 sec
on 200 mV range
Input Impedance
10 MΩ, 11 MΩ on 6 V range
10 MΩ
400 mV: >100 MΩ, 4V: 10 MΩ
40 V - 1000 V: 9.1 MΩ
10 MΩ