Table 16 - Edge Trigger Setup Menu
Menu
Option
Setting
Description
Coupling
DC
Passes all components of the signal
AC
Blocks DC components and attenuates
signals below 170 Hz.
HF Reject
Attenuates the high-frequency
components above 140 kHz.
LF Reject
Blocks the DC component and attenuates
the low-frequency components below 7
kHz.
Hold off
Adjusts the hold off time.
Hold off
Reset
Reset hold off time to 100 ns.
Return
Return the Edge Trigger Menu.
Hold Off: To adjust the hold off time, go to the Edge Trigger Setup Menu
and press
, then use the
or
arrow keys to adjust
the hold off time, which will be displayed when it is adjusted.
Hold off time can be reset to 100 ns by pressing
to select Hold Off
Reset.