Edge Trigger
Edge trigger seeks to trigger off of a specified slope and level of a
waveform.
Select Edge under Type from the trigger menu.
Table 15 - Edge Trigger Menu
Menu
Option
Setting
Description
Type
Edge
Trigger on the rising or falling edge of the input
signal.
Source
CH1
CH2
Set CH1 or CH2 as the trigger source.
Slope
Trigger on rising edge of the trigger signal.
Trigger on falling edge of the trigger signal.
Trigger on rising and falling edge of the trigger
signal.
Mode
Auto
Use this mode to let the acquisition free-run in
the absence of a valid trigger; This mode allows
an un-triggered, scanning waveform at 100
ms/div or slower time base settings.
Normal
Use this mode when you want to see the
triggered waveforms only; the scope will not
acquire waveform until it’s been triggered.
Single
The setup detects a trigger and acquires
waveform, then stops to give a signal capture.
Set Up
Enter the Edge Trigger Setup Menu.