beautypg.com

B&K Precision 2712 - Datasheet User Manual

True rms ac + dc tool kit, Rue rms, Model 2712 data sheet

background image

w w w.bk prec ision.c om

Tel.: 714.921.9095

10 A range: ±(2.0% rdg + 10 dgts)

Input Protection

0.5 A/500 V & 10 A/500 V fast blow ceramic fuses

10A Input

10 A for 60 sec max. followed by a 10 minute cooling period

Voltage Burden

40 mA, 10 A ranges: 0.2V; 400 mA range: 1 V

AC Current (True RMS) (50 Hz-1 kHz)

Ranges

40 mA, 400 mA, 10 A

Resolution

1 uA, 10 uA, 10 mA

Accuracy

40 mA to 400 mA ranges: ±(2.0% rdg + 30 dgts)
10 A range: ±(2.5% rdg + 30 dgts)

Crest Factor

≤ 3

AC coupled True RMS

Specified from 10% to 100% of range

Voltage Burden

40 mA, 10 A ranges: 0.2V; 400 mA range: 1 V

Input Protection

0.5 A/500 V & 10 A/500 V fast blow ceramic fuses

10A Input

10 A for 60 sec max. followed by a 10 minute cooling period

Resistance

Ranges

400 Ω, 4 kΩ, 40 kΩ, 400 kΩ, 4 MΩ, 40 MΩ

Resolution

0.01 Ω, 0.1 Ω, 1.0 Ω, 10 Ω, 1 kΩ, 10 kΩ

Accuracy

400 Ω range: ±(0.3% rdg + 15dgts)
4 kΩ to 400 kΩ ranges: ±(0.3% rdg + 5dgts)
4 MΩ range: ±(0.5% rdg +10 dgts)
40 MΩ range: ±(1.5% rdg + 20 dgts)

Open Circuit Voltage (typical) 1.2 Vdc (2.5 Vdc on 400 Ω range)
Overload Protection

500 VDC/ACrms

Capacitance

Ranges

4 nF, 40 nF, 400 nF, 4 uF, 40 uF

Resolution

1 pF, 10 pF, 100 pF, 1 nF, 10 nF

Accuracy

4 nF range: ±(3.0% rdg + 20 dgts)
40 nF to 400 nF ranges: ±(3.0% rdg + 5 dgts)
4 uF to 20 uF ±(3.0% rdg + 5 dgts)
20 uF to 40 uF ±(5.0% rdg + 5 dgts)

Overload Protection

500 VDC/ACrms

Frequency

Ranges

100 Hz, 1 kHz, 10 kHz, 100 kHz, 500 kHz

Resolution

0.01 Hz, 0.1 Hz, 1 Hz, 10 Hz, 100 Hz

Accuracy

±(0.1 % rdg + 10 dgts)

Sensitivity

5 Hz-100 kHz: >500 mVrms, 100 kHz-500 kHz: >1.5 Vrms

Minimum Pulse Width

>2 us

Duty Cycle Limits

>30 % and <70 %

Overload Protection

500 VDC/AC rms

Diode Test

Test Current

0.8 mA (typical)

Resolution

0.1 mV

Accuracy

±(1.5 % rdg + 10 dgts)

Open Circuit Voltage

3.0 VDC (typical)

Overload Protection

500 VDC/ACrms

Continuity

Audible Indication

Less than 40 Ω

Response Time

500 ms

Overload Protection

500 VDC/ACrms

General

Display

4 3/4 digit, 40000 count LCD

Analog Bargraph

40 segments with measurements of 20 times per second

Polarity

Automatic, positive implied, negative polarity indication

Overrange

MSD (Most Significant Digit) Blinks

Low Battery Indication

Battery symbol will display when battery voltage drops below
operating levels

Measurement Rate

2 times per second

Auto Power Off

Approximately 30 minutes

Operating Environment

0ºC to 50ºC at <70 % relative humidity

Storage Environment

-20ºC to 60ºC at 0 to 80 % relative humidity

Accuracies

Stated accuracy at 23ºC +5ºC at <75 % relative humidity

Temperature Coefficient

0.1 x (specified accuracy) per ºC. (ºC to 18ºC, 28ºC to 50ºC)

Power

Single standard 9 volt battery, NEDA 1604, JIS 006P, IEC 6F22

Battery Life

150 hours typical

Dimensions

165 mm (H) x 78 mm (W) x 42.5 mm (D)

Weight

Approximately 10 oz. (285 g) with battery and hanger

True RMS AC + DC Tool Kit

®

DMM

Model 2712

Data Sheet

Technical data subject to change
© B&K Precision Corp. 2013
Page 1 of 2

The 2712 is a full featured auto ranging true RMS DMM that combines
performance, value and functionality. The 40000 count LCD has a green back
light that allows the meter to be used in dimly lit environments. The 2712 can
measure resistance values up to 40 Mohm and capacitance to 40 uF. With the
auto power off feature you won’t have to worry about accidental battery
consumption. Value-packed features make this meter a must in every "Tool Kit

®

".

DC Volts

Ranges

400 mV, 4 V, 40 V, 400 V, 1000 V

Resolution

0.01 mV, 1 mV, 10 mV, 100 mV, 1 V

Accuracy

400 mV range: ±(0.15 % rdg + 10 dgts)
4 V to 1000 V ranges: ±(0.1 % rdg + 5 dgts)

Input Impedance

400 mV: >3 MΩ; 4 V to 1000 V: 2.3 MΩ

Overload Protection

1000 VDC or 750 VACrms

AC Volts (True RMS) (45 Hz-1 kHz)

Ranges

400 mv, 4 V, 40 V, 400 V, 750 V

Resolution

0.01 mV, 1 mV, 10 mV, 100 mV, 1 V

Accuracy

400 mV to 400 V ranges: ±(1.2% rdg + 20dgts) 45 ~ 60 Hz
4 V range: ±(1.5% rdg + 20 dgts) 60 ~ 500 Hz
40 V to 400 V ranges: ±(1.5% rdg + 20dgts) 60 ~ 1 kHz
750 V range: ±(2.0% rdg + 20 dgts) 45 ~ 500 Hz

AC + DC Volts (True RMS) (45 Hz-1 kHz)

Ranges

400 mv, 4 V, 40 V, 400 V, 750 V

Resolution

0.01 mV, 1 mV, 10 mV, 100 mV, 1 V

Accuracy

400 mV to 400 V ranges: ±(1.5% rdg + 20dgts) 45 ~ 60 Hz
4 V range: ±(2.0% rdg + 20 dgts) 60 ~ 500 Hz
40 V to 400 V ranges: ±(2.0% rdg + 20dgts) 60 ~ 1 kHz
750 V range: ±(2.0% rdg + 20 dgts) 45 ~ 500 Hz

Crest Factor

≤ 3

AC coupled True RMS

Specified from 10% to 100% of range

Input Impedance

400 mV: >3 MΩ; 4 V to 750 V: 2.3 MΩ

Overload Protection

1000 VDC or 750 VACrms

DC Current

Ranges

40 mA, 400 mA, 10 A

Resolution

1 uA, 10 uA, 10 mA

Accuracy

40 mA to 400 mA ranges: ±(0.5% rdg + 10 dgts)

Specifications

2712

model

Supplied Accessories: User manual, one set of test leads, one spare fuse, and 9V battery (installed)

Three-Year Warranty

RUE RMS

T

Select between Auto or
Manual ranging

A ut o/ Manual Ranging