8 Specifications
Oscilloscope Specifications
All specifications apply to measurements with a 10X probe. To verify that
the oscilloscope meets specifications, the oscilloscope must first meet the
following conditions:
The oscilloscope must have been operating continuously for 30
minutes within the specified operating temperature.
If the operating temperature changes by more than 5° C, you must
perform the Do Self Cal operation, accessible through the Utility
menu.
All specifications are guaranteed unless noted “typical.”
Models
2511
2512
2515
2516
Bandwidth
60 MHz
100 MHz
60 MHz
100 MHz
Real Time
Sampling Rate
1 GSa/s (half-channel interleaved)
(1)(2)
, 500 MSa/s (per
channel)
Channels
2 (non-isolated)
2 (isolated)
Rise Time
< 5.8 ns
< 3.5 ns
< 5.8 ns
< 3.5 ns
Ch to Ch
Isolation (Both
channels in
same V/div
setting)
> 100:1 at 50 MHz
Memory Depth
40 kpts (half-channel interleaved)
(1)(2)
, 20 kpts (per
channel)
Deep Memory
(3)
2 Mpts (half-channel interleaved)
(1)
, 1 Mpts (per channel)
Vertical
Resolution
8 bits
Vertical
Sensitivity
2 mV/div – 100 V/div (1-2-5
order)
5 mV/div – 100 V/div (1-2-5
order)
DC Gain
Accuracy
5 mV/div-100 V/div: ≤ ± 3
%
5 mV/div-100 V/div: ≤ ± 3
%