If any conditions occur that could be considered as equipment failure or
malfunction, follow the inspection procedures given below.

CHAPTER 9

T

ROUBLESHOOTING

G

UIDE

Check

item

Problem

Details of inspection

Corrective action

Reference

page

Measure-
ment
value
display

• Measured value is not

displayed.

• Nothing appears on the

display.

• Is the power switch ON?

• Is the AC power cable con-

nected?

• Turn ON the switch.

• Connect the AC power cable.

p. 5

"------" is displayed.

• Are targets in position?
• Is the entire scanning area

completely covered?

• Is the appropriate measuring

segment selected?

• Is the scanning head properly

connected?

• Is there any dust or dirt on

the laser emitting/receiving
window?

• Is the laser emission indicator

LED lamp lit?

• Wasn’t "------" displayed

immediately after turning ON
the power switch?

• Wasn’t "------" displayed

immediately after switching
the measurement modes?

• Wasn’t "------" displayed

immediately after shorting
the laser emission remote
control terminal?

• Isn’t the number of measure-

ments for averaging set to a
large value?

• Is there condensation forming

on the equipment?

• Correctly place the targets

within the effective measure-
ment area.

• Correctly select the appropri-

ate measuring segment.

• Turn the power switch OFF,

then after connecting the
scanning head, turn the
power switch ON

• Remove dust or dirt from the

laser emitting/receiving
window.

• Short LASER REMOTE

terminals.

• Wait approximately 10

seconds after turning ON the
power switch before opera-
tion.

• Wait approximately 3

seconds after changing
measurement modes before
resuming operation.

• Wait approximately 10

seconds after shorting this
terminal.

• In the Average mode, maxi-

mum of approximately 2.6
seconds are required until
the measurement value is
displayed after the target is
positioned in the measure-
ment area.

• Change the location.

p. 84

p. 28

p. 15

p. 21