Chapter 8 error messages, H-err, P-err – KEYENCE LS-3100/3100W Series User Manual
Page 84: Calib error, Chapter 8, Error messages, H-err p-err calib error, Rror, Essages

78
CHAPTER 8
E
RROR
M
ESSAGES
H-Err
• If no scanning head is
connected to the controller or the connected scanning head malfunctions,
an error message appears on the main display (CH1).
When 1 scanning head is used:
If no scanning head is connected or the connected scanning head mal
functions, "H-Err" appears.
When 2 scanning heads are used:
When "HEAD2 ON" is selected using PRM:
If no scanning head is connected to the HEAD1 connector port or the
scanning head connected to that port malfunctions, "H-Err" appears.
If no scanning head is connected to the HEAD2 connector port or the
scanning head connected to that port malfunctions, "H2-Err" appears.
When "HEAD2 OFF" is selected using PRM:
Since the LS controller judges that no scanning head is to be connected
to the HEAD2 connector port, "H-Err" appears.
• When "H-Err" or "H2-Err" appears, turn OFF the power. Then, make sure
that the problem with the scanning head is removed or that the scanning
head is connected properly, then turn ON the power.
• Turn OFF the power to cancel error message "H-Err" or "H2-Err".
P-Err
When the transmitter and receiver are not aligned in terms of the optical
axis, "P-Err" appears on the main display (CH1). Align them correctly.
Note: Ignore this message when using model LS-3033 SO.
When the LS controller is not calibrated correctly, "CALIB ERROR" appears
on the main display (CH1). Carefully read the description for the calibration
procedures, and calibrate the LS controller again using a proper gauge.
CALIB ERROR
08.10.1, 10:57 AM
78