KEYENCE CB-EP100 User Manual
Page 69

3-4 Message Communication
3-45
3
Communi
cat
in
g wit
h the
L
J
-V700
0 Ser
ies
- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -
Instance ID = 1
Data unique to a measuring instrument (LJ-V) connected to the CB-EP100 can be
obtained. The setting values obtained are the settings of the setting area currently in
operation for the program number at that point in time.
Service
code
(hexadecimal)
Attribute
ID
(hexadecimal)
Name
Description and service data
Attribute
Data type
0EH
65H
Setting values are
being collected
Indicates whether or not the latest set-
ting information of measuring instru-
ment is currently being collected. 0
indicates that the information is not
being collected.
R
DWORD
66H
Save memory
status
Indicates whether data is being writ-
ten to the save memory. 0 indicates
that data is not being written.
DWORD
67H
Error code
Stores the error codes of connected a
measuring instrument.
UINT
68H
Warning code
Stores the warning codes of con-
nected a measuring instrument.
UINT
69H
-
6DH
Reserved
4DH
6EH
Trigger
For information on detailed formats,
refer to
by the instance service 4DH (control
input)" (Page 3-78).
-
6FH
Auto-zero
70H
Timing
71H
Reset
72H
Memory clear
73H
Batch start
74H
Batch end
75H
Storage start
76H
Storage end
78H
-
2FFH
Reserved
0EH
300H
System Status
For information on detailed formats,
refer to
(312 Bytes)) Assembly Instance
(Instance ID): 64H" (Page 3-9)
R
WORD
301H
Measuring in
strument status
302H
Valid status
303H
Judgment
standby status
304H
Alarm status
305H
-
307H
Reserved