- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -
3-4 Message Communication
3-32
3
Communi
cat
in
g wit
h the
L
J
-V700
0 Ser
ies
(8) Initializing settings
Initialize the settings for the specified program number specified in the setting
write area in a measuring instrument connected to the CB-EP100.
"(8) Initializing settings" (Page 3-82)
(9) Confirming profile data
Save the profile data in a measuring instrument connected to the CB-EP100 so that
it can be read.
"(9) Confirming profile data" (Page 3-83)
(10)Setting the date and time
Update the date and time in a measuring instrument connected to the CB-EP100
to the specified date and time.
"(10) Setting the date and time" (Page 3-84)
Command: 00H
Response
CB-EP100
Scanner
Service code: 50H
Class ID: 6BH
Instance ID: 01H
Attribute ID: None
Service data: 00H
Measuring
instrument
The setting write area of program
number 0 is initialized.
Command
Response
CB-EP100
Scanner
Service code: 57H
Class ID: 6BH
Instance ID: 01H
Attribute ID: None
Service data: None
Measuring
instrument
The latest profile is saved.
Command:
Service data
Response
CB-EP100
Scanner
Service code: 58H
Class ID: 6BH
Instance ID: 01H
Attribute ID: None
Service data:
UDINT: Number of data bytes
UDINT: Year (last two digits)
UDINT: Month
UDINT: Day
UDINT: Hour
UDINT: Minute
UDINT: Second
Measuring
instrument
The date and time are