Statistics > include/exclude sample (command), Statistics > include/exclude value (command), O_4133 – BUCHI DuMaster D-480 User Manual
Page 333: O_332, O_4134
Operation Manual D-480
©BÜCH Labortechnik AG
10 - Menu and dialog descriptions
333
Statistics > Include/Exclude sample (Command)
What is the command used for?
This command either includes a sample in the calculation or excludes it.
Short cuts
Invocation via the toolbar:
Invocation via hot key: CTRL + T
Note
You can execute this command in the following cases:
In the statistics view: The selected sample of a statistical group is displayed crossed-out
(disabled) and the statistics calculation is performed without this sample. If you execute the
command again, the sample is displayed as "enabled" again and the statistical calculation is
performed with this sample again. If only two samples are enabled in a statistical group, you
cannot disable either of these samples.
In the calibration view: The selected sample of a calibration series is displayed as a full circle
(eliminated) and is excluded from the calibration calculation. Executing this command again
displays the sample as "uneliminated" and includes it in the calibration calculation again.
Statistics > Include/Exclude value (Command)
What is the command used for?
This command either includes the value of a sample in the calculation or excludes it.
Note
You can execute this command in the following cases:
In the statistics view: The selected value of a sample of a statistical group is displayed crossed-
out (disabled) and the statistics calculation is performed without this value. If you execute the
command again, the value is displayed as "enabled" again and the statistical calculation is
performed with this value again. If only two values are enabled in a statistical group, you cannot
disable either of these values.