7 - Using the instrument

146

 Allocate the daily factor to the real samples in the sample view (text view). Enter the daily factor

in the corresponding factor column.

How to selectively trigger the daily factor

Proceed as follows:

 Define the type of factor calculation. For that purpose, open the "Configure Calculations" dialog in

the operating software by selecting Options > Settings > Calculation.

 Define the mode of factor determination:

If you in section "Factor determination"...

... select "sequentially" then the daily factor is computed group-by-group and changes
throughout the day.

... select "entirely" the daily factor is computed on the basis of all factor samples in the file.
An average factor is computed.

 Then click OK to save your settings.

 Measure a standard substance, that must be defined in Options > Settings > Standard

Samples, as a factor sample.

 If you only want to determine the daily factor from a certain number of samples, select these

samples in the sample view (text view).

 Open the "Daily factors" dialog in the operating software by selecting Math. > Factor.

 If you selected a certain number of samples for the calculation, check "Use standard samples

only".

 Click Yes to continue. The daily factor is computed automatically and allocated to the subsequent

samples.

How to set automatic daily factor determination

Proceed as follows:

 Define the type and degree of automation for factor calculation. For that purpose, open the

"Configure Calculations" dialog in the operating software by selecting Options > Settings >
Calculation
.

 Define the mode of factor determination:

If you in section "Factor determination"...

... select "sequentially" then the daily factor is computed group-by-group and changes
throughout the day.

... select "entirely" the daily factor is computed on the basis of all factor samples in the file.
An average factor is computed.

 Specify the degree of automation for factor determination. Check "Automatic factor

determination". factor calculation.

 Then click OK to save your settings. As soon as you measure a substance defined as a factor

sample, the daily factors are automatically computed afterwards.

Recommendation for automated daily factor determination

When automatically computing the daily factor, it is useful to enable the "Standard tolerance
monitoring" function. This checks standard samples for fluctuations of results. If the tolerance define
in Options > Settings > Standards is exceeded, continuous analysis is stopped.

Enabling "Standard tolerance monitoring"

Proceed as follows:

 Open the "Error handling" dialog in the operating software by selecting Options > Settings >

Error handling.

 Check "Standard tolerance monitoring".