Standard samples (dialog), H_859, H_846 – BUCHI DuMaster D-480 User Manual
Page 308: O_408

Operation Manual D-480
©BÜCH Labortechnik AG
10 - Menu and dialog descriptions
308
Label
Element type
Meaning
autom. Auto-
increment the number
in the sample name
Radiobutton
If a sample name ends with a number (e.g. "Name001"),
this figure is automatically incremented by a certain amount
and suggested for the next sample (e.g. "Name002").
Exceptions:
If the number of characters is completely used up (e.g.
"Name999"), it is not incremented and the same name
is suggested for the next sample.
If the figure is not at the end of the sample name (e.g.
"Name001xyz"), it is not incremented.
autom. Auto-
decrement the
number in the sample
name
Radiobutton
Works the same way as auto-increment, only the figure in
the name is decremented.
Exception:
The figure "zero" is not decremented.
Increment/decrement
interval
Textbox with
Buttons
In this field you can input the interval at which to
increment/decrement the figure in the name. This figure
can be between 1 and 99. The number must, however, fit
in the displayable number of digits, otherwise the number is
not changed.
Fill method names to
stop marker
Checkbox
If, for example, you have input/copied 10 successive
sample names with the sample memory empty, the stop
marker is now at the tenth sample. If you have input a
method name for the first sample, this name is entered for
all ten samples.
Fill daily factors to
last sample
Checkbox
The daily factor input for the first sample is entered for all
subsequent samples.
Fill blank values to
last sample
Checkbox
The blank value input for the first sample is entered for all
subsequent samples.
OK
Button
Saves new settings
Cancel
Button
Discards settings not yet saved.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Settings for sample input (on page 133).
Optimizing sample data editing (on page 99).
Standard samples (Dialog)
Function
In this dialog you can create, delete, view, copy and change standard samples.
Open
Open the dialog by selecting Options > Settings > Standards.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
---
Listbox
This field lists all standard samples already defined.