Error buffer (dialog), Error display (dialog), O_404 – BUCHI DuMaster D-480 User Manual
Page 306: O_405
Operation Manual D-480
©BÜCH Labortechnik AG
10 - Menu and dialog descriptions
306
Remaining elements:
Label
Element type
Meaning
Close
Button
Closes the dialog.
Help
Button
Opens the description of the dialog in the help.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Performing a system test.
Error buffer (Dialog)
Function
This dialog displays a chronological list of all errors and events detected by the system. Every event
detected is accompanied by the following information for tracking purposes:
Date and time in the "Date/Time" column.
Error code in the "Event" column.
Open
Open the dialog by selecting Options > Diagnostics > Error buffer.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
Date/time; event
Listbox
Field listing every event with the date, time and its error
code.
Number of events
Label
Field displaying the total number of events currently in the
error buffer.
show deleted items
Checkbox
If activated, all events which have been deleted previously
will be displayed again. This element is not enabled at the
lowest access level.
Delete
Button
Deletes selected events.
Button
Prints the content of the error buffer.
Close
Button
Closes the dialog.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Interpreting PC error messages (on page 240).
Error display (Dialog)
Function
This dialog decodes the various error codes and displays the results. The error types are divided into
different categories and each category is allocated to a separate tab.