BUCHI DuMaster D-480 User Manual
Page 115

Operation Manual D-480
©BÜCH Labortechnik AG
6 - Starting up and shutting down the instrument
115
CO
2
intake pressure
CO
2
intake pressure is set at the delivery point so that the following values are displayed on screen:
The "Input" display shows approx. 1300 mbar.
The "MFC CO2" display shows approx. 650 ml/min.
The "Flow CO2" display shows approx. 650 ml/min.
Flow rate
The following flow rate settings are typical:
The "MFC CO2" display shows approx. 650 ml/min for the measuring gas outlet.
The values for the reference gas outlet are between 30ml/min and 60ml/min. There is no display
for the reference gas outlet.
During standby, the "MFC O2" display shows 0 ml/min. The corresponding value will be
displayed only during O
2
dosing (combustion).