BUCHI DuMaster D-480 User Manual
Page 312

Operation Manual D-480
©BÜCH Labortechnik AG
10 - Menu and dialog descriptions
312
If you are using multiple analyzers with one PC, you must specify one instrument identifier on the
"Identifier" tab so that the LIMS data can be allocated to the correct instrument. You can also
define user names.
Note
Note that the LIMS port name must not be identical to the port name between the PC and the
analyzer. This could lead to a system crash.
Open
Open the dialog via Options > Settings > LIMS.
Elements
The following describes the elements of the dialog.
"LIMS data" tab
In section "Data to send":
Label
Element type
Meaning
{Data member}
Checkbox
Sends the corresponding data member.
Send data
automatically after
every analysis
Checkbox
Automatically send the data to LIMS after every analysis.
Send weighing data
on input/transfer
Checkbox
After entering or transferring the sample weight, it will be
sent automatically to the LIMS.
Note: Sends every element with a check mark. The order corresponds to the list from top to bottom. If
you change the selection of elements to send, you must edit the LIMS software, otherwise the data
may be misinterpreted.
"Communication" tab:
In section "Data transfer":
Label
Element type
Meaning
none
Radiobutton
No data are sent to LIMS.
via serial port
Radiobutton
Data are transferred via the serial port.
via network
Radiobutton
The data to send to the LIMS are written to a file.
The file can be accessed from the LIMS and from the
instrument operating software.
In section "Serial":
Label
Element type
Meaning
Port name
Textbox
Field to input the name of the interface.
The port name must not be the same as the port for
communication between the PC and analyzer. Otherwise,
operation of the analysis system will be disrupted.
Baud rate
Textbox
Field to input transfer parameters.
Databits
Textbox
Field to input transfer parameters.
Stopbits
Textbox
Field to input transfer parameters.
Parity
Textbox
Field to input transfer parameters.
Note: The textboxes in the "Serial" section only have to be filled in if you have selected the "via serial
port" radio button in the "Data transfer" section. The settings here must match those of the LIMS
interface.