System > auto zero (command), System > wake up (command), Feeding (dialog) – BUCHI DuMaster D-480 User Manual
Page 291: O_371, O_376, O_377, 291 system > wake up (command)

Operation Manual D-480
©BÜCH Labortechnik AG
10 - Menu and dialog descriptions
291
Note
The "Abort analysis" dialog appears.
Observe the notes listed therein. Abort analysis only if either the sample has not been thrown in yet or
has been already completely combusted. In the latter case, make sure that the column does not
retain any element components, i.e. the desorption procedure has been completed. Otherwise, an
abortion leads to system contaminations combined with significant damages.
To abort a running analysis, you have to confirm the note displayed in the dialog first. Only after the
confirmation the Abort analysis button becomes accessible, thus allowing the abortion.
System > Auto zero (Command)
What is the command used for?
This command causes to set the current output values of the dectors to zero. The purpose is to reset
the signals to zero if they have drifted over time.
System > Wake up (Command)
What is the command used for?
This command wakes up the "sleeping" system. The carrier gas is reactivated and the furnaces are
heated back up to their working temperatures. This command is used to manually wake up the
system if you need to start up the system prematurely, i.e. earlier than the chosen wake-up time.
Short cuts
Invocation via the toolbar:
Note
If you have defined a wake-up time that is in the past, you must wake up the sleeping system again
with this command.
Feeding (Dialog)
Function
In this dialog you specify the size of the sample carousel. Make sure that the setting here matches to
the actual size of the sample carousel.
Open
Open the dialog by selecting System > Feeding.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
Carousel with {n}
positions
Radiobutton
The feeding system is switched to a carousel size of {n}
positions.
OK
Button
Saves the setting.