BUCHI DuMaster D-480 User Manual
Page 32

Operation Manual D-480
©BÜCH Labortechnik AG
3 - Product description
32
Interfaces
USB, RS232 /V24, Safety low voltage as per DIN IEC 380/VDE
0806/08.81
Dimensions and weights
The following table contains dimensions and weight of the instrument:
Reference value
Technical specifications
Dimensions
480 mm x 550 mm x 550 mm (Width x Depth x Height)
Weight
65 kg
Electrical and gas supply
The following table contains the technical specifications of the electrical and gas supply:
Reference value
Technical specifications
Mains voltage
100–230 Volt AC ± 10 %; 50–60 Hz
Power requirement
0.6 kWh/h
Connected load
1.4 kVA
Connected electrical loads and
frequency of the complete
analysis unit
Connected electrical load:
Voltage:
Frequency:
16 A
100V AC ± 10 %
120V AC ± 10 %
230V AC ± 10 %
50–60 Hz
Supply gases
Carbon dioxide:
Oxygen:
Purity: 99.995 %
Purity: 99.995 %
Consumption of supply gases
Per analysis approx. 4 litres carbon dioxide.
Per analysis approx. 0.1 to 1.0 litres oxygen.
Ambient conditions
The following table contains the technical specifications of the ambient conditions:
Reference value
Technical specifications
Protection class
Protection class I, protective conductor connection
Mode of protection
IP20, installation in dry rooms only
Noise level
55 dB (A), as per DIN 45635 section 1
Permissible ambient temperature
Maximum: + 35 °C, minimum: + 15 °C