Analysis mode (dialog), H_782, O_3766 – BUCHI DuMaster D-480 User Manual
Page 293
Operation Manual D-480
©BÜCH Labortechnik AG
10 - Menu and dialog descriptions
293
Label
Element type
Meaning
Position
Radiobutton
Moves the carousel moves to the desired position. You can
set the position number in the textbox to the right. You
must input the desired position number minus 1.
Reference run
Radiobutton
The carousel performs reference run, i.e. the carousel first
moves to a defined reference position and then to position
0 which has almost the same location.
OK
Button
Adjusts the position of the carousel, if confirmed above that
all samples have been removed from the carousel.
Cancel
Button
Aborts the process.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Performing measurement work (on page 136).
Reacting to a power failure (on page 248).
Performing calibration.
Analysis mode (Dialog)
Function
In this dialog you can switch the operating software to a new operating mode. Before switching
operating modes, it may be necessary to refit the analyzer. After refitting you can switch operating
mode.
Caution
Overheating if tube fillings are not appropriate for the operating mode!
Overheated tube fillings melt, run into the furnace area and destroy the furnace.
Make sure that the tube fillings correspond to the selected operating mode.
Open
Open the dialog via System > Mode. First a message window is displayed indicating that the
analyzer may need to be refitted before switching operating modes. Only if you have this message
confirmed by Yes the "Analysis mode" dialog will appear.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
{Mode}
Radiobutton
The operating software is switched to the corresponding
mode.
Solids
Radiobutton
The instrument software is changed over to solids mode.
Manual liquid
injection
Radiobutton
The instrument software is changed over to liquid mode
with manual injection.
VLS liquid sampler
Radiobutton
The instrument software is changed over to liquid mode
with VLS sample feeder.
OK
Button
Switches the operating mode.
Cancel
Button
Aborts the process.