beautypg.com

BUCHI DuMaster D-480 User Manual

Page 183

background image

Operation Manual D-480

©BÜCH Labortechnik AG

8 - Maintaining the instrument

183

The following table explains the characteristics relating to the sample, as displayed in the above
picture:

Characteristic

Description

Sample No.

Shows the analysis number of the sample selected at the cursor.

Area

Integral of the selected sample.

slope

Slope of the curve at the calibration point of the selected sample.

{Element name}

theo

Theoretical absolute content of the selected sample

{Element name}

act

Absolute content of the selected sample, computed with the current curve

Diff.

Difference between "theo" and "act", absolute

Diff.

Difference between "theo" and "act", relative

Aids for assessing the calibration curve

The following aids are available for assessing the calibration curve:

Via View > Zoom in / Zoom out you can display the graph so that you can visually check the
curve.

You can click on the circle to select any sample. This displays the characteristics of the sample.

You can also select a sample using the buttons

and

. This displays the

characteristics of the sample.

Assessing the calibration curve

The following are used to assess the calibration curve:

Process standard deviation "Proc-SD"

Quality factor "r"

Coefficient "a", that corresponds to a calibrated blank value.

Process standard deviation "Proc-SD"

Use the following values for process standard deviation "Proc-SD" as guidelines:

Acceptable value

Proc-SD = ± 3 µg in the lower range
Proc-SD = ± 15 µg in the upper range

Very good value

Proc-SD = ± 1 µg

Quality factor "r"

Use the following values for quality factor "r" as guidelines:

Good value for linear calibration function

r = 0.99 (r should tend to 1)

Good value for polynomial as calibration
function

r = < 2-4

Coefficient "a"

Use the following values for coefficient "a" as guidelines:

Acceptable range

a = 1 x 10

-3

to 3 x 10

-2

or smaller

The range given here applies to calibration with one calibration range and, when calibrating with two
calibration ranges, only for the lower range, while "a" may be greater in the second range.

If you do not approximately achieve the listed values you can assume that:

either you have selected the wrong calibration curve calculation mode,

or the calibration measurement has failed (weighing error, etc.).