9 - Repairing the instrument
246
The electrical section is now visible.
Pull out the fuse holder of the appropriate fuse.
Press the fuse out.
Check the fuse:
If the fuse wire...
... is visible: Check whether the wire is broken or not.
... is not visible: Measure fuse continuity.
Make a decision for the further proceeding:
If...
... the fuse wire is broken: Insert a new fuse of the same type.
... no continuity was measured on the fuse: Insert a new fuse of the same type.
... the fuse wire is OK: Insert the old fuse again.
... continuity was measured on the fuse: Insert the old fuse again.
Phase 3: Insert fuses
Proceed as follows:
Make sure that the fuse specifications match the label on the fuse holder.
Insert the fuse into the appropriate fuse holder and tighten the semiconductor relay holding plate
again.