O_3663, O_4312 – BUCHI DuMaster D-480 User Manual
Page 113

Operation Manual D-480
©BÜCH Labortechnik AG
6 - Starting up and shutting down the instrument
113
Removing the furnace transport protection
Proceed as follows:
Open the front door of the instrument.
Push the right side door open from inside (magnetic lock). The transport protection is now visible
in the form of a bracket (1).
Unscrew the screws and remove the bracket.
Fasten the bracket on the rear casing angle (1). It should be stored there so that it does not get
lost.
Close the right side door and screw the big screw back in that keeps the door closed.
Removing the water trap transport protection
Proceed as follows:
Open the front door of the instrument.