Chapter 9
Maintenance
Cleaning
Clean the Bird Site Analyzer only with a soft cloth dampened with mild detergent
and water. Do not use any other type of cleaning solution.
Charging the Battery
The internal battery pack will automatically recharge when the Site Analyzer is
powered from the AC or cigarette lighter adapter. Recharging time, from a full
discharge, is approximately 4 hours.
External Battery Pack (Optional), Long-term Storage
Maintenance
When the external battery pack is stored for long periods of time without
proper maintenance, it will lose its charge and also lose its ability to reach full
charge when put into service.
The maintenance procedure is to apply a one-hour “top off” charge at three
month intervals. Follow the battery charge procedures on page 7 for one hour
only. Do not over charge the battery in storage.
Note: If a battery has been stored for an extended period of time
without “top off” charging, perform a full charge followed by a full dis-
charge. Repeat this cycle if necessary up to three times. If the battery
does not remain charged after three charge-discharge cycles, replace
the battery.
CAUTION
Harsh or abrasive detergents, and some solvents, can damage the display unit
and labels.