Viewing a Recalled Trace

When a saved trace is recalled, it is displayed along with the trace currently being
measured (two traces on the screen). With a current and recalled trace, the dis-
play behaves as follows:

The frequency range will change to the settings of the recalled trace.

Recalled traces adopt the display type of the current trace (envelope,
maximum, minimum).

The resolution of the recalled trace does not affect the resolution of
the current trace.

Markers always follow the current trace, not the recalled trace.

When comparing the current trace to a recalled trace, each data point on
the current trace is compared to only the first data point on the recalled
trace (at the frequency of the marker).

The status box at the top of the screen displays the resolution and dis-
play type of the current trace on the top line and the label text of the
recalled trace on the bottom line.

Note: Calibration is automatically turned off if the recalled trace has
a different frequency range from the trace currently being measured.

To remove a recalled trace from the display:
1.

Press the Save/Recall softkey.

2.

Press the Esc key.